Tornier, Inc recalls TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX18MM NON STERILE, catalog number DWJ318
- Recall date
- December 19, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1237-2026
- FDA classification
- Class II
- Brand / firm
- Tornier, Inc
- Sold / distributed
- US Nationwide distribution.
Why it was recalled
A specific lot of screws is incorrectly labeled. Screws are labeled as 5.0mmx18mm but actual screws are 5.00mmx14mm. If mismatch is detected during surgery, the issue may cause a minor delay in surgery. If not detected and incorrect screw is used, issue may increase risk of earlier mechanical failure, loosening, and need for revision surgery.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX18MM NON STERILE, catalog number DWJ318
Get recall alerts
Free email alert whenever Tornier, Inc has a new recall — straight from official government data. Unsubscribe anytime.