Siemens Healthcare Diagnostics Inc recalls epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;
- Recall date
- April 2, 2026
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2005-2026
- FDA classification
- Class II
- Brand / firm
- Siemens Healthcare Diagnostics Inc
- Sold / distributed
- Worldwide distribution - US Nationwide and the countries of Italy, Spain, Austria, Portugal, Ireland, France, Romania, Canada, India, Philippines, Mexico.
Why it was recalled
Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading can lead to inappropriate treatment of nonexistent acidosis (respiratory of metabolic) or missed diagnosis of alkalosis (a true high pH: respiratory or metabolic alkalosis), potentially causing harm through unnecessary interventions with the former or failure to deliver appropriate treatment with the latter.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;
Get recall alerts
Free email alert whenever Siemens Healthcare Diagnostics Inc has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Siemens Healthcare Diagnostics Inc