Medical device recalls Moderate risk

Philips North America recalls Philips Ingenia 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781342. 2. Model Number (REF): 781377. 3. Model…

Recall date
April 14, 2026
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1958-2026
FDA classification
Class II
Brand / firm
Philips North America
Sold / distributed
Worldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar, Netherla…

Why it was recalled

The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Philips Ingenia 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781342. 2. Model Number (REF): 781377. 3. Model Number (REF): 782103. 510(k) Numbers: K163116, K173079, K183063, K193215, K213516, K213583, K223458, K251397, K251808

Get recall alerts

Free email alert whenever Philips North America has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Philips North America