Medical device recalls

Heraeus Medical GmbH (Dental Division) recalls Heraeus, PALAMIX uno. Material Number: 66057893.

Recall date
March 25, 2026
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1997-2026
FDA classification
Class III
Brand / firm
Heraeus Medical GmbH (Dental Division)
Sold / distributed
Worldwide distribution - US Nationwide and the countries of Aland Islands, Algeria, Australia, Austria, Belgium, Brazil , Chile, China, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Kosovo, Lithuania, Luxembourg, Malaysia, Malta,…

Why it was recalled

Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).

Class III means the product violates regulations but is not likely to cause harm.

What was recalled

Heraeus, PALAMIX uno. Material Number: 66057893.