GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH Nuclear Medicine System
- Recall date
- June 20, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2098-2025
- FDA classification
- Class II
- Brand / firm
- GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
- Sold / distributed
- Worldwide - US Nationwide distribution.
Why it was recalled
Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH Nuclear Medicine System