Medical device recalls High risk

Boston Scientific Corporation recalls Boston Scientific Cardiac Resynchronization Therapy (CRT)labeled as: 1. VALITUDE CRT-P EL, Model Number U125; 2. VALITU…

Recall date
March 19, 2026
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1771-2026
FDA classification
Class I
Brand / firm
Boston Scientific Corporation
Sold / distributed
worldwide

Why it was recalled

Updated software, Brady software maintenance release 6 (SMR6), is now available for the ACCOLADE" family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps)1 and the advisory population is expanding to include all CRT-P and dual-chamber extended life (DR-EL) devices. Dual chamber modes are specifically indicated for treatment of the following: " Conduction disorders that require restoration of AV synchrony, including varying degrees of AV block " VVI intolerance (i.e., pacemaker syndrome) in the presence of persistent sinus rhythm " Low cardiac output or congestive heart failure secondary to bradycardia

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

Boston Scientific Cardiac Resynchronization Therapy (CRT)labeled as: 1. VALITUDE CRT-P EL, Model Number U125; 2. VALITUDE CRT-P EL MRI, Model Number U128; 3. VISIONIST CRT-P EL, Model Number U225; 4. VISIONIST CRT-P EL, Model Number U226; 5. VISIONIST CRT-P EL MRI, Model Number U228.

Get recall alerts

Free email alert whenever Boston Scientific Corporation has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Boston Scientific Corporation