FILMARRAY 2 recalled over Salmonella risk
- Recall date
- August 7, 2024
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- BioFire Diagnostics, LLC recalls FILMARRAY 2.0 INSTRUMENT, REF: FLM2-ASY-0001, FLM2-ASY-0001-W
- Recall number
- Z-3122-2024
- FDA classification
- Class II
- Brand / firm
- BioFire Diagnostics, LLC
- Sold / distributed
- Worldwide - US Nationwide distribution including in the states of NC, TX, OH, CA, KS, AL, WA, NY, PA, IL, WI, KY, LA, NE, FL, UT, IA, SC, VA, AZ, CO, WV, MO, TN, MN, DC, DE, OK, WY, IN, NJ, NM, ID, OR, MS, SD, MI, MT, GA, HI, NV, MD, MA, AR, NH, PR, ND, VT, CT, AK, ME, RI and the countries of Korea…
Why it was recalled
Some automated in vitro diagnostic modules and instruments manufactured or serviced between May 2022 and May 2024 were calibrated with internal equipment that may lead modules and instruments to report higher melting temperatures, which may lead to erroneous results including false positive Salmonella, Giardia lamblia, false positive/negative Norovirus, and false negative MRSA.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
FILMARRAY 2.0 INSTRUMENT, REF: FLM2-ASY-0001, FLM2-ASY-0001-W
Get recall alerts
Free email alert whenever BioFire Diagnostics, LLC has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: BioFire Diagnostics, LLC